Abstract

Background

Even though walking can be sustained for great distances, considerable energy is required for plantarflexion around the instant of opposite leg heel contact. Different groups attempted to reduce metabolic cost with exoskeletons but none could achieve a reduction beyond the level of walking without exoskeleton, possibly because there is no consensus on the optimal actuation timing. The main research question of our study was whether it is possible to obtain a higher reduction in metabolic cost by tuning the actuation timing.

Methodology/Principal Findings

We measured metabolic cost by means of respiratory gas analysis. Test subjects walked with a simple pneumatic exoskeleton that assists plantarflexion with different actuation timings. We found that the exoskeleton can reduce metabolic cost by 0.18±0.06 W kg−1 or 6±2% (standard error of the mean) (p = 0.019) below the cost of walking without exoskeleton if actuation starts just before opposite leg heel contact.

Conclusions/Significance

The optimum timing that we found concurs with the prediction from a mathematical model of walking. While the present exoskeleton was not ambulant, measurements of joint kinetics reveal that the required power could be recycled from knee extension deceleration work that occurs naturally during walking. This demonstrates that it is theoretically possible to build future ambulant exoskeletons that reduce metabolic cost, without power supply restrictions.

Comments

© 2013 Malcolm et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Journal Title

PLoS ONE

Volume

8

Issue

2

Included in

Biomechanics Commons

Share

COinS